SDI-12 Depth Sensor
Short Description
The pressure transducers of the series MPSDI are designed to measure
pressure and water level with high accuracy (better 0.01ft.). The transducers
are available either as submersible
unit connected to a vented
cable for error-free compensation of barometric pressure changes or
as thread
mount type used e.g. in connection with a bubbler
system. The incorporated electronics compensates nonlinearity and also
temperature variations in a wide range resulting in an increased overall
accuracy.
The SDI-12 compatible serial-digital interface enables output of pressure,
level and temperature values as well as the user-specific programming of
the sensors in a multidrop bus confuguration. Via the SDI-12-interface,
MPSDI pressure sensors can be connected to computers, data loggers etc.
by means of a shielded three wire cable Recalibration, free adjustement
of offset and measuring-span (scaling) and choice of output-parameters
are possible. Low power circuitiy combined with a sleep-mode keeps power
consumption down to an absolute minimum.
Advantages

- SDI-12 serial-digital output
- Microprocessor-controlled transducer
- High accuracy (0.05 %)
- Low power consumption
- Temperature compensated
- Environmentally rugged
- High reliability and stability
Applications

- For water well measurements
- Level sensors for small-bore applications
- Measurement of gas pressures
- Pressure / level measurement instorage and run-of-river power stations,
water supply plants, sewage treatment plants
- Level mesurements in hydrology (navigation locks, weirs, irrigation
plants etc.)
Technical Characteristics

|
|
| Measuring ranges |
0 ... 0.1 bar to 0 ... 50 bar
(0 ... 145 psi up to 0 ... 725 psi) |
| Overpressure |
3 to 6 times (depending on the range) |
| Accuracy |
< 0.05% of full scale or 3 mm (0.01 ft) for ranges < 5 m (15
ft) |
| Temperature influence |
typ. < 30 ppm/ºC of full scale
max. < 50 ppm/ºC of full scale |
| Operating temperature |
-20ºC ... +70ºC (-4ºF ... 158ºF)
-38ºC ... +70ºC (-36.4ºF ... 158ºF) optional |
| Communication |
Bidirectional serial bus, multidrop communication, up to 10 sensors,
SDI-12 protocol |
| Overvoltage protection |
Built-in protector up to 2000 V (transient) |
| Control and Programming |
Recalibration, free adjustement and choice of output-parameters via
data logger or a personal computer (in connection with a SDI-12 to RS232
converter) |
|